JOURNAL ARTICLE
New electrode material GCE/AgNPs‐D3 as an electrochemical sensor used for the detection of thallium ions.
Published In: Electroanalysis, 2023, v. 35, n. 5. P. 1 1 of 3
Database: Academic Search Ultimate 2 of 3
Authored By: Modrzejewska‐Sikorska, Anna; Konował, Emilia; Karbowska, Bożena; Szatkowska, Dominika 3 of 3
Abstract
Optimal working parameters for the innovative electrode GCE/AgNPs‐D3 were determined and the selectivity was assessed. The stripping anodic peak current of thallium characterized in linearity over range from 4.9 ⋅ 10−8 to 4.9 ⋅ 10−7 mol ⋅ dm−3, with LOD = 7.16 ppb (3.5 ⋅ 10−8 mol ⋅ dm−3). Furthermore, the new electrode showed more than sevenfold improvement in the thallium current signal compared to the unmodified electrode GCE. High selectivity was achieved using EDTA disodium salt as a complexing agent for other metals. [ABSTRACT FROM AUTHOR]
Additional Information
- Source:Electroanalysis. 2023/05, Vol. 35, Issue 5, p1
- Document Type:Article
- Subject Area:Geology
- Publication Date:2023
- ISSN:1040-0397
- DOI:10.1002/elan.202200281
- Accession Number:163765413
- Copyright Statement:Copyright of Electroanalysis is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Looking to go deeper into this topic? Look for more articles on EBSCOhost.