Monte Carlo simulation of shielding designs for a cabinet form factor preclinical MV‐energy photon FLASH radiotherapy system.
Published In: Medical Physics, 2023, v. 50, n. 5. P. 3055 1 of 3
Database: Academic Search Ultimate 2 of 3
Authored By: Rosenstrom, Andrew; Santana‐Leitner, Mario; Rokni, Sayed H.; Shumail, Muhammad; Tantawi, Sami; Dewji, Shaheen; Loo, Billy W. 3 of 3
Abstract
Purpose: A preclinical MV‐energy photon FLASH radiotherapy system is being designed at Stanford and SLAC National Accelerator Laboratory. Because of the higher energy and dose rate compared to conventional kV‐energy photon laboratory‐scale irradiators, adequate shielding in a stand‐alone cabinet form factor is more challenging to achieve. We present a Monte Carlo simulation of multilayered shielding for a compact self‐shielding system without the need for a radiation therapy vault. Methods: A multilayered shielding approach using multiple alternating layers of high‐Z and low‐Z materials is applied to the self‐shielded cabinet to effectively mitigate the secondary radiation produced and to allow the device to be housed in a Controlled Radiation Area outside of a radiation vault. The multilayered shielding approach takes advantage of the properties of high‐Z and low‐Z radiation shielding materials such as density, cross‐section, atomic number of the shielding elements, and products of radiation interactions within each layer. The Monte Carlo radiation transport code, FLUKA, is used to simulate the total effective dose produced by the operation. Results: The multilayered shielding designs proposed and simulated produced effective dose rates significantly lower than monolayer designs with the same total material thickness at the regulatory boundary; this is accomplished through the manipulation of the locations where secondary radiation is produced and reactions due to material properties such as neutron back reflection in hydrogen. Borated polyethylene at 5 wt% significantly increased the shielding performance as compared to regular polyethylene, with the magnitude of the reduction depending upon the order of the shielding material. Conclusions: The multilayered shielding provides a path for shielding preclinical FLASH systems that deliver MV‐energy bremsstrahlung photons. This approach promises to be more efficient with respect to the shielding material mass and space claim as compared to shielded vaults typically required for clinical radiation therapy with MV photons. [ABSTRACT FROM AUTHOR]
Additional Information
- Source:Medical Physics. 2023/05, Vol. 50, Issue 5, p3055
- Document Type:Article
- Subject Area:Physics
- Publication Date:2023
- ISSN:0094-2405
- DOI:10.1002/mp.16290
- Accession Number:163605204
- Copyright Statement:Copyright of Medical Physics is the property of Wiley-Blackwell and its content may not be copied or emailed to multiple sites without the copyright holder's express written permission. Additionally, content may not be used with any artificial intelligence tools or machine learning technologies. However, users may print, download, or email articles for individual use. This abstract may be abridged. No warranty is given about the accuracy of the copy. Users should refer to the original published version of the material for the full abstract. (Copyright applies to all Abstracts.)
Looking to go deeper into this topic? Look for more articles on EBSCOhost.